Maintenance
206
9. If the window is dirty clean it by following the procedure
10. If the window is damaged, replace with a new radial purge window (Long Radial
Purge Window B0810377 or Short Radial Purge Window N0690672).
11. Return the new radial purge window in the stainless steel ring to the instrument.
12. Carefully screw it into place.
13. Use the tool to tighten down the stainless steel ring and radial purge window into
place.
14. Return the Shear Gas Nozzle. See
Returning the Shear Gas Nozzle
15. Turn on the plasma.
Cleaning the Windows
1. Rinse the windows thoroughly with deionized water.
2. To remove more stubborn marks:
Soak the windows in warm, 40°C, 10-20% nitric acid for one hour.
Rinse the windows thoroughly with deionized water.
3. Dry them with an optical-quality cloth.
4. If the windows are scratched or badly etched, discard them and install new ones.
Note
Do
not
rub the surfaces of the windows before you have cleaned them as
described below. The more you rub the surface of the windows, the more
chance there is of scratching them. You must discard scratched windows.
Allow to dry completely before reinstalling. If there is any moisture on the
inside of the window, it can degrade the UV performance and/or damage
the optics.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......