Maintenance
209
3. See the following procedure on installing a new shear gas nozzle or replacing the
existing shear gas nozzle.
Returning the Shear Gas Nozzle
Use this procedure to install a new shear gas nozzle or returning the existing shear
gas nozzle if you had to remove it in order to replace either the axial or radial purge
window.
1. Place the shear gas assembly back into position. Insert the two screws into the
slots on the bracket. Make sure the screws are located at the top of the slot before
tightening with the T20 Torx screwdriver (
2. Reinstall the torch assembly; see the section
earlier in this chapter.
3. Align the torch and perform an axial and radial alignment; refer to the procedure
Setting the Torch Viewing Position
Note
Before you perform an analysis at wavelengths below 190 nm, purge the
optics overnight on high purge. For other analyses, purge the system until
you obtain a steady signal.
Table 5-4.
Replacement Parts: Purge Components
Part No.
Description
09921036
O-Ring (for Dual View Radial Purge Tube)
09921057
Purge Outlet O-Ring
09992731
Axial Purge Window
N0690672
Short Radial Purge Window
B0810377
Long Radial Purge Window
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......