14
Environment
Operating Conditions
Pollution Degree
This equipment will operate safely in environments that contain non-conductive
foreign matter up to Pollution Degree 2 in EN 61010–1.
Recommended operating conditions:
•
Indoors.
•
The location must be free of smoke, dust, corrosive fumes, direct sunlight and
excessive vibration.
•
Ambient tempe15
C to +35
C (+59
F to +95
F), with a maximum
change not exceeding 2.8
C (5
F) per hour. . For optimum instrument
performance, the room temperature should be controlled at 20° ± 2 °C. F
•
Relative humidity 20% to 80%, without condensation.
•
Altitude in the range -400 m to 2,000 m (-1,312 ft. to 6,562 ft).
Safe operating conditions:
•
Indoors
•
Tempe5 °C to +40 °C (+41 °F to +104 °F).
•
Relative humidity 20% to 80%, without condensation.
•
Altitude in the range -400 m to 2,000 m (-1,312 ft. to 6,562 ft).
Storage conditions:
•
Ambient temperature –20 °C to +60 °C (–4 °F to +140 °F).
•
Relative humidity 20% to 80%, without condensation.
•
Altitude in the range -400 m to 12,000 m (-1,312 to 39,370 feet).
Explosive Atmosphere
This instrument is not designed for operation in an explosive atmosphere.
Avertissement : atmosphère explosive
Cet instrument n'est pas conçu pour fonctionner dans une atmosphère
explosive.
!
Warning
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......