Maintenance
184
Cleaning the Torch
The torch components should be cleaned regularly to remove accumulated deposits.
After prolonged use, a torch may lose its transparency and become crystalline. In this
condition, called devitrification, the torch is very fragile. You should not attempt to
clean a devitrified torch, but it can still be used. If the torch is cloudy, this generally
does not affect performance because the plasma is either viewed through the slot on
the torch, or, for axial viewing, through the open end of the torch.
If you have been analyzing aqueous solutions, the torch glassware and injector tube
should be cleaned in an acid bath. An ultrasonic bath may be used. Start with a
solution of 5% nitric acid or aqua regia and, if deposits persist, increase the acid
concentration, up to 20% if necessary. Acid solutions must be used with care.
Observe all of the manufacturer's safety recommendations. These recommendations
are normally provided in a material safety data sheet (MSDS) supplied with the
chemical.
Use of hydrofluoric acid for cleaning the torch is
not
recommended.
If you have been analyzing organic solutions, the torch glassware and injector tube
can be cleaned using a solvent or diluted soap solution. Be sure to inspect the torch
and injector daily for carbon buildup. A muffle furnace can be used if necessary to
remove organic deposits on the torch or injector. Place the torch or injector in a
muffle furnace at 500
C to 550
C for several hours.
After cleaning the torch, dry it thoroughly using clean air or nitrogen. No moisture
should be present.
Caution
Attention
To extend the life and aid in preventing devitrification of glassware, it is
necessary to remove all traces of alkali on glassware caused by handling it
with bare hands. To remove the alkali, wipe the glassware clean with
cotton wool dipped in alcohol.
Pour prolonger la durée de vie et de l'aide dans la prévention de la
dévitrification de la verrerie, il est nécessaire d'éliminer toute trace d'alcali
sur la verrerie provoquée par la manipulation à mains nues. Pour supprimer
l'alcali, essuyez la verrerie propre avec un coton imbibé d'alcool.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......