Maintenance
253
5. Tighten the adjustment screw
slowly
(turn clockwise) until droplets just start
forming at the end of the SAMPLE OUT side of the sample tubing.
6. Tighten the adjustment screw until you see a smooth flow of bubbles (3 turns on
the peristaltic pump). Add an additional three more turns.
7. Turn off the pump. The tension is now correctly set. No bubbles should be seen
in the sample tubing once the sample tube is kept in the solution.
Cleaning the Pump
Periodically inspect the peristaltic pump rollers to make sure they are clean and move
freely. Clean the exterior of the pump with a cloth moistened with water.
The pump head on the peristaltic pump can be removed and cleaned if necessary. A
replacement pump head is available (Part No. 09220266).
Removing the Pump Head on the Peristaltic Pump
To remove the pump head on the peristaltic pump:
1. Remove the pump tubing.
2. Unscrew the large screw in the center of the pump head. Place the screw and
washer in a secure place.
3. Lift the pump head off. To clean the pump head, use water or a mild solvent and
dry thoroughly.
4. Return the pump head to the pump. Orient the pump head by aligning the notch
on the bottom of the pump head with the pin on the shaft. The pump head must
click in this slot for a proper fit. Replace the washer and tighten the screw.
Caution
Attention
If a spill occurs on the pump, immediately clean it up.
A corrosive spill
could damage the pump.
Si un déversement se produit sur la pompe, nettoyer immédiatement
vers le haut.
Un déversement corrosif pourrait endommager la pompe.
Содержание AVIO 200
Страница 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...Safety Practices 1 ...
Страница 32: ...30 Safety Practices ...
Страница 56: ...54 Safety Practices ...
Страница 57: ...Preparing Your Laboratory 2 ...
Страница 58: ...Preparing Your Laboratory 56 ...
Страница 70: ...Preparing Your Laboratory 68 ...
Страница 71: ...System Description 3 ...
Страница 72: ......
Страница 84: ...System Description 82 Figure 3 7 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view ...
Страница 97: ...Installation 4 ...
Страница 98: ......
Страница 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Страница 164: ...Installation 162 ...
Страница 165: ...Maintenance 5 ...
Страница 166: ......
Страница 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Страница 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Страница 272: ...Maintenance 270 ...
Страница 273: ...Troubleshooting 6 ...
Страница 274: ......
Страница 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Страница 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Страница 296: ...Troubleshooting 294 Figure 6 4 Aux Gas too high Air Leak or Spray Chamber Temperature too high ...
Страница 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Страница 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Страница 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Страница 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Страница 308: ......