D. FilterTec Plus Specific issues
When this occurs:
Check the following:
Possible Solution
You have a balance in use,
but aren’t getting any data
from the balance.
Is it connected and
configured correctly?
Review suggestions for the
“Scale Error” problem listed
above. If your not getting this
error, see #2 that follows.
Did you turn the Filtrate
Weight Alarm off?
If the Filtrate Weight Alarm
under Alarm Enable: Filtrate
Wgt. Is set to OFF, the unit will
not look for the balance, and not
gather data from it.
The lack of “Scale Initialization”
occurring is a good clue. Set
this alarm to either Alarm Only,
or Pump Stop. If you don’t want
the alarm to occur, set the Limit
to either 0.0 or well beyond
anything you might collect.
The pressure reading is
always 0.0, even when the
sensor is disconnected.
This occurs when the
sensor is Zeroed when not
connected, or Span is
pressed instead of Zero,
thereby setting the max
input to a zero value.
You must perform a Factory
Reset to correct this.
Following the Reset, verify the
Motor RPM setting, and be sure
to only Zero the sensors when
they are connected!
Do not use the Span button
unless the Pressure Sensor
inputs are being calibrated.