46
An Ohaus Adventurer Pro AV8101 scale is normally sent with the system when
purchased. If a different scale is required, please contact Parker Technical Support for
configuration information.
5.11 Ohaus Adventurer Pro Balance Parameters:
Press and hold the Menu button until MENU appears on the display. Release this button, and
now use the Yes, No, and Back buttons to navigate the Sub-menus. Press the Tare button to
exit the menu.
Please set the following Menu Items, all others are left at their factory defaults:
ReadOut
AutoZero Off
Filter
Low
Print-1
Output
Whenstable Off
AutoPrint Off
Content
Num Only
Off
Header
Off
Gross
Off
Net Off
Tare
Off
Reference Off
Result
On
GLP
Off
Layout
Line Format
Multi
RS232-1
Baud
9600
Parity
7 No Parity
Handshake Off
In the Setup Mode, Scale Manuf., select “OHAUS3”. By making this selection, the FilterTec Plus
will implement the correct parameters for communicating with the Adventurer Pro Series
balance. You will also need Parker P/N 080-067PGS, Balance Cable. (Note: The 080-066
Parker Ohaus Balance cable will not work with this balance series.)
NOTE:
The Scale Communication settings for the updated (have the blue buttons with white
text) FilterTec Plus and FilterTec Plus Plus have returned to match the settings used by all other
Parker bench top systems. They no longer require continuous output.