4.0 PROGRAMMABLE R-P: How to generate
a PROGRAMMABLE R-P program.
SUMMARY:
In PROGRAMMABLE R-P Mode you can generate a program that changes flow
rates or controlled pressure rates in a stair-stepped (STEP) or ramped (SCAN) manner over
time. If you wish, you can also control a pair of rotary selector valves, or repeat the program
automatically.
When generating or editing a Programmable R-P program, all program statements that you
want to see implemented during a specific timing block or interval
must precede the TIME
statement for that block.
For example, on the next page, the program statement: 001 CW, 002
RUN, and 003 RATE are all implemented at the beginning of the first timing block defined by
program statement 004 TIME: 00:05 (five minutes).
CW, Pump Direction:
This program statement is implemented by pressing the front panel key
with the double arrows. This key functions like a toggle switch and allows you to select either
CW (clockwise) or CCW (counter-clockwise) pump direction.
RUN:
This statement is implemented by pressing the front panel key labeled “RUN”, it instructs
the FilterTec Plus to turn on the motor.
TIME, Timing Interval:
Press TIME key, then select the desired timing interval in
hours:minutes. All preceding program statements are implemented at the beginning of this TIME
statement.
STOP:
This statement is implemented by pressing the “STOP” key on the front panel. It
instructs the FilterTec Plus to turn off the motor.
RATE, R-Step, R-Scan:
Press “RATE” key then select RATE by pressing the “A” button. Select
R-Step to set a flow rate to be maintained for a given Time period. Select R-Scan to set a flow
rate to be scanned (ramped) up or down to over a given Time period. Please be sure that the
selected flow rate does not exceed the capacity of the installed pump head /motor combination.
PRESSURE, P-Step, P-Scan:
Press “RATE” key then select PRESSURE by pressing the “C”
button. Select P-Step to set a pressure to be maintained for a given Time period. Select P-Scan
to set a pressure to be scanned (ramped) up or down to over a given Time period. The
controlled pressure is the one specified in SETUP, PUMP, Press. Sensor, Source.
Editing the program is accomplished using the “NEXT” “DELETE” and “LAST” keys. NEXT
moves to the next line in the program, LAST moves to the previous line in the program, (i.e.
backspace) and DELETE removes the line of the program currently displayed. To remove and
replace a command, press “DELETE” and “LAST”, and then input the new command.