When this occurs:
Check the following:
Possible Solution
When “Scale Error” “Hit any
key” shows on your screen.
Is your scale turned on, and
are the cables tight?
Press any key to clear the error,
tighten the cables, turn on the
scale, and try it again.
Ok, the scale is on, cables
are ok, and it still won’t
work?
Refer to Section 5, Setup: Scale
of this manual, then go to Setup
Mode of the pump, select Scale,
then Scale Mfr, and confirm the
selection is correct.
Ok, the scale is chosen
correctly, and it still doesn’t
work. Now what?
Either refer to the same manual
section mentioned above to
check the scale settings, or call
Parker tech support for help
correcting them if you are not
using the default scale.
When the pump is acting
weird. The flows and
pressures are all wrong; it
gets data from the balance,
but slows down way to
early; etc.
Has someone messed with
your FilterTec Plus?
If someone other than yourself
or your supervisor has changed
the settings without your
knowledge, you can to return
them to their original settings.
Have you had electrical
problems in the building
lately?
Power spikes and brown outs
can cause problems. Enter
Setup; Pump, and then select
Factory Reset. This step is a
last resort. This will return the
unit to the factory default values.
You will need to return to Setup;
Pump; Motor RPM, and verify its
setting, as well as Setup, Pump;
Pump Head to verify it as well.
Call Parker tech support if
needed.