FilterTec Plus System Specifications:
Mechanical:
Dimensions:
Width: 5.75 in (14.6cm); Height: 8.5 in (212.6); Depth: 11in (27.9)
Weight:
14 lbs (6.4kg)
Enclosure:
Aluminum / Steel; Corrosion Resistant, Recessible Handle
Pump Head / Motor Options:
Three Masterflex Easy-Load II (77201-60) adjustable occlusion pump heads with 8
RPM motor
, accommodates thin-walled (0.060”) pump tubing: #13, 14, 16, 25, and #17.
Pump Rate Range: 0.03 to 24.3 ml/min.
Three Masterflex Easy-Load II (77201-60) adjustable occlusion pump heads with 160-
RPM motor,
accommodates thin-walled (0.060”) pump tubing: #13, 14, 16, 25, and #17.
Pump Rate Range: 0.5 to 554 ml/min.
Three Masterflex Easy-Load II (77201-62) adjustable occlusion pump heads with 600-
RPM motor,
accommodates thick-walled (0.090”) pump tubing: #15, 24 and #35. Pump
Rate Range: 59 to 2,258 ml/min.
Pressure Sensors:
Accommodates up to three (3) SciPres ™ Disposable pressure
sensors: P1, P2, and P3. Pressure control is achieved by modulating pump output. The
user selects P1, P2 or P3 as the controlled pressure.
Pressure Displayed
with a resolution of 0.1 psi; choice of
bar, psi, kpa
.
Pressure Range:
The default pressure range of the sensors is
0-60psi
, and is calibrated at
the factory. Most peristaltic pumps generate pressures up to 40 psi. If you have need for
higher pressures, contact Parker Customer Service for assistance.
Electrical
:
Power:
90 - 264 V
, 47-63 Hz, 150 VA, listed Class 2 switching power supply; double
fused: 1A-T, 250V (CE: IR35A 250V
).
Battery:
CR1220, used to support the internal clock only, not user serviceable.
Motor:
Choice of three (3) motors: 8, 160 and 600 RPM at 30V
, 3.8 Amperes, Variable
Pump Speed optically encoded, servo-controlled motors.
Encoder:
100 lines/rev for 600-RPM motor. 120 lines per/rev for 8 & 160-RPM motors.
I/O Ports:
“
Printer
”, Female DB9 connector for data collection with Printer or PC.
“
S1
”, “
S2
”, “
S3
”, Male DB9 connector for RS-232 connections to an electronic scale.
“
External I/O
”, Female DB37 connector used for remote On/Off control of FilterTec Plus via
footswitch, or for Analog interface with 4-20 ma sources, A1, A2, A3.
“
V
”, Female DB15 connector, Not utilized on FilterTec Plus.
“
Temperature
”, 2 pin Conxall connector for SciTemp™ disposable Temperature Sensor.
“
P1, P2, P3
”, RJ11 connectors used for SciPres disposable Pressure Sensors.
“
USB
”, USB-A connector, used for RS-232 data collection with a PC.
“
Ethernet
”, RJ-45 connector, not enabled.
Display:
Two-line LCD, 20 characters each, back-lit.
Data Entry
: Membrane keyboard with auditory feedback.
Environmental
:
Temperature range: 4 to 40
C.
Altitude: up to 2000 Meters
Indoor, dry environments only, clean-up is wipe down only.(IP 20)
Relative humidity: 0-95%
Voltage fluctu/- 10%
Pollution degree: 2