Setting the Output Conditions
90
ZW User's Manual
Setting the Median Filter
The intermediate value of multiple sets of data can be output as the measurement result.
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Filter]
Item
Setting item
Setting value
Description
Medial filter
Filter mode
OFF
(default value)
Median filter is not used.
LOW
Outputs the intermediate value of the last three measurements.
Middle
Outputs the intermediate value of the latest nine measurements.
HIGH
Outputs the intermediate value of the last 15 measurements.
1
Select how many of the latest intermediate
values to output from [Medial filter] - [Filter
mode].
The median filter can also be set by the operating keys on the Sensor Controller.
Setting the Median Filter p.236