Error Messages
ZW User's Manual
209
8
Tro
uble
s
hootin
g
Errors Common to All Communication States
These errors occur in common regardless of communication state. When these errors occur, the ALARM
output of parallel I/O turns ON, and "SYSERR" and error code "CODE.XX" are displayed on the main and sub-
displays, respectively.
If an error code other than the one listed above is displayed, the displacement sensor may be broken. Please
contact an OMRON branch or sales office.
*1:
As a provisional measure, the measurement can be resumed using the data of the previously read Calibration ROM.
<Operation method>
With error code 3 displayed on the sub-display, hold down the Mode switching key, then when [OK/CAN] is displayed, press the ZERO/SET
key.
<Cautions>
- When using this method, always check the serial number of the previously read Calibration ROM in "controller information," then check
that it matches the Sensor Head side serial number. Measurement will not be correct unless they match.
- When restarting the main unit, perform the same operations again.
This operation is disabled for a displacement sensor into which no Calibration ROM has ever been inserted and started up.
Error Code
Name
Description
Cause
Remedy
02
Linearity correction
data error
The linearity correction
data is corrupted.
Calibration ROM damage Check to make sure that the
Calibration ROM is correctly
inserted. If correctly inserted, the
Calibration ROM or displacement
sensor may be broken.
Please contact an OMRON branch
or sales office.
(*1)
03
Linearity correction
data read error
Reading of the linearity
correction data was not
executed correctly.
Calibration ROM not
inserted
07
System setting error
The system settings
saved to the Sensor
Controller are corrupt.
The displacement sensor
power was turned OFF
during saving/loading of
system settings.
After holding down the Mode
switching key, press the ZERO/SET
key to clear the system settings and
the bank data, then resume the
starting process.
08
Bank data error
The bank data saved to
the Sensor Controller is
corrupt.
The displacement sensor
power was turned OFF
during saving/loading of
bank data.
11
Firmware update
error
Failed to update the
firmware by
WarpEngineZW.
WarpEngineZW was
interrupted during an
update.
Do not operate WarpEngineZW
during an update.
If this error occurs, turn the
displacement sensor ON again and
retry the update.
25
Ethernet
communication
parameter error
An invalid IP address is
set.
Invalid IP address setting
Change to the correct IP address.
Содержание ZW-C1*T
Страница 17: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 16 1 2 Basic Operation Flow 17 ...
Страница 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Страница 116: ...Setting the System 114 ZW User s Manual MEMO ...
Страница 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Страница 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Страница 319: ...Index ZW User s Manual 317 Index 11 ...