Specifications and External Dimensions
294
ZW User's Manual
Sensor controller
Specifications
Item
Specifications
ZW-C10T
ZW-C10AT
ZW-C15T
ZW-C15AT
Input/Output type
NPN
PNP
Number of connected Sensor Heads
1 per Sensor Controller
Sensor Head compatibility
Available
Light source for measurement
White LED
Segment
display
Main display
11-segment red display, 6 digits
Sub-display
11-segment green display, 6 digits
LED display
Status indicators
HIGH (orange), PASS (green), LOW (orange), STABILITY (green), ZERO
(green), ENABLE (green), THRESHOLD-H (orange), THRESHOLD-L
(orange), RUN (green)
External inter-
face
Ethernet
1 port, 100BASE-TX/10BASE-T
RS-232C
1 port, 115,200 bps max.
20-pole ter-
minal block
Judgment output
(HIGH 1/PASS 1/LOW 1)
Transistor output system
Output voltage: 21.6 to 30 VDC
Load current: 50 mA or less
Residual voltage when turning ON: 1.2 V or less
Leakage voltage when turning OFF: 0.1 mA or less
BUSY output (BUSY 1)
ALARM output (ALARM)
ENABLE output
(ENABLE 1)
Analog voltage output
(OUT1V)
-10 to +10 V, Output impedance: 100
Analog current output
(OUT1A)
4 to 20 mA, Max. load resistance: 300
LED OFF input
(LED OFF 1)
DC input system
Input voltage: 24 VDC ± 10% (21.6 to 26.4 VDC)
Input current: 7 mA Type. (24 VDC)
ON voltage/ON current: 19 V/3 mA or less
ON voltage/ON current: 5 V/1 mA or less
ZERO RESET input
(ZERO 1)
TIMING input (TIMING
1)
RESET input (RESET 1)
52-pole
extension
connector
Binary
Measured
value output
(BINARY 0 to
20)
Transistor output system
Output voltage: 21.6 to 30 VDC
Load current: 50 mA or less
Residual voltage when turning ON: 1.2 V or less
Leakage voltage when turning OFF: 0.1 mA or less
Gate signal
output (GATE)
Selected task
output
(BINARY_OU
T1/2)
Selected task
input
(BINARY_SE
L1/2)
DC input system
Input voltage: 24 VDC ±10 % (21.6 to 26.4 VDC)
Input current: 7 mA Typ. (24 VDC)
Voltage/Current when turning ON: 19 V/3 mA or more
Voltage/Current when turning OFF: 5 V/1 mA or less
Содержание ZW-C1*T
Страница 17: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 16 1 2 Basic Operation Flow 17 ...
Страница 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Страница 116: ...Setting the System 114 ZW User s Manual MEMO ...
Страница 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Страница 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Страница 319: ...Index ZW User s Manual 317 Index 11 ...
Страница 321: ......