Settings for Bank Control
This section describes the settings for controlling banks by using parallel I/O.
Selecting banks
The bank is selected in combinations of the bank select input signals (BANK_SEL1 to 3).
• Bank switching is begun 0.2 seconds after the input state changes.
• At most it takes about 100ms to switch banks.
• During bank switching, the BUSY output becomes ON.
• If the bank mode is set to [JUDGE], the bank cannot be switched at the external signal input because the number of
banks increases to 32.
Outputting the currently selected bank number
The currently selected bank number is output.
The output bank number depends on the combination of the bank number output signals (BANK_OUT1 to 3).
Bank selection input 1
(BANK_SEL1)
Bank selection input 2
(BANK_SEL2)
Bank selection input 3
(BANK_SEL3)
Selected bank
OFF
OFF
OFF
BANK1
ON
OFF
OFF
BANK2
OFF
ON
OFF
BANK3
ON
ON
OFF
BANK4
OFF
OFF
ON
BANK5
ON
OFF
ON
BANK6
OFF
ON
ON
BANK7
ON
ON
ON
BANK8
Bank number output 1
(BANK_OUT1)
Bank number output 2
(BANK_OUT2)
Bank number output 3
(BANK_OUT3)
Output bank
OFF
OFF
OFF
BANK1
ON
OFF
OFF
BANK2
OFF
ON
OFF
BANK3
ON
ON
OFF
BANK4
OFF
OFF
ON
BANK5
ON
OFF
ON
BANK6
OFF
ON
ON
BANK7
ON
ON
ON
BANK8
Important
Содержание ZW-C1*T
Страница 17: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 16 1 2 Basic Operation Flow 17 ...
Страница 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Страница 116: ...Setting the System 114 ZW User s Manual MEMO ...
Страница 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Страница 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Страница 319: ...Index ZW User s Manual 317 Index 11 ...
Страница 321: ......