Setting Measurement Items
234
ZW User's Manual
3
Sets the surface peak hold for TASK1 and the surface average hold for TASK2.
4
Press the
/
keys to select "TASK3" and press the
key.
5
Press the
/
keys to select "ITEM" and press the
key.
6
Press the
/
keys to select "CALC" and press the
key.
7
Select "PARAM.K" for the parameter type.
Press the
/
keys to select "PARAM.K" and press the
key.
8
Set the parameter to "0".
The current setting value is displayed on the sub-display.
Press
key to enter the editing mode, and the sub-display
blinks.
To edit numerical values, use the
/
/
/
keys.
Input "0", then press the
key.
9
The decimal point is displayed.
Press the
/
key to move the decimal point.
Determine the decimal point and then press the
key.
10
Press the
key to return to the previous menu.
11
Select "PARAMm" for the parameter type.
Press the
/
keys to select "PARAM.M" and press the
key.
12
Set the parameter to "1".
The current setting value is displayed on the sub-display.
Press
key to enter the editing mode, and the sub-display
blinks.
To edit numerical values, use the
/
/
/
keys.
Input "1", then press the
key.
13
Press the
key to return to the previous menu.
Steps
Key operation
Display
Description
Содержание ZW-C1*T
Страница 17: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 16 1 2 Basic Operation Flow 17 ...
Страница 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Страница 116: ...Setting the System 114 ZW User s Manual MEMO ...
Страница 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Страница 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Страница 319: ...Index ZW User s Manual 317 Index 11 ...
Страница 321: ......