Setting the Output Conditions
88
ZW User's Manual
Performing Thickness Scaling
The thickness is measured at one position and offset values are set for that measured value.
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Scaling]
Item
Setting item
Setting value
Description
Thickness scaling
Setting value
-999.999999 to 999.999999 [mm]
Correct the measured value, and set the value to display.
1
Select scaling ON/OFF from [Scaling].
2
Click [Auto setting].
The [Scaling] popup menu appears.
3
Select [Thickness scaling], and click [Next
>].
The [Scaling] popup menu display changes to
[Thickness settings].
4
Click [STOP] to fix the [Current value], set
[Set value] and click [Next >].
The [Scaling] popup menu display changes to
[Confirm the scaling parameters], and the span
and offset values are displayed.
5
Click [OK].
Thickness scaling can also be executed by the operating keys on the Sensor Controller.
Performing Thickness Scaling p.243
The current measured value is displayed at
[Current value], and is refreshed at 100 ms
cycles. When [STOP] is clicked, refreshing of the
[Current value] display stops. Clicking the
[STOP] button again resumes display refreshing.
Note
Note
Содержание ZW-C1*T
Страница 17: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 16 1 2 Basic Operation Flow 17 ...
Страница 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Страница 116: ...Setting the System 114 ZW User s Manual MEMO ...
Страница 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Страница 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Страница 319: ...Index ZW User s Manual 317 Index 11 ...
Страница 321: ......