Wiring
40
ZW User's Manual
2-4 Wiring
20-pole terminal block
Used for judgment output, control input, etc.
Compatible cable specifications: AWG 18 to 28, pin processed length: 7 mm
• Wire the power supply separate from other devices. Wiring together with other devices or using the same duct
could cause a false operation or damage the sensor.
• Do not turn off the power during initial processing immediately after the Sensor Controller power is turned on, as the
memory inside the Sensor Controller is being accessed. This may corrupt the data.
• Cut the unnecessary signal lines so that they do not touch other signal lines.
• The length of Cables should be less than 30 m.
Signal name
No.
Signal name
OUT1 (V)
1
11
NC
OUT1 (A)
2
12
NC
OUT1 0V
3
13
NC
COM_OUT1
4
14
ALARM
HIGH1
5
15
BUSY1
PASS1
6
16
ENABLE1
LOW1
7
17
ZERO1
COM_IN1
8
18
RESET1
24 VDC (+)
9
19
TIMING1
24 VDC (-)
10
20
LED OFF1