Setting the Output Conditions
104
ZW User's Manual
Setting the Zero Reset Memory
Select whether or not to hold the measured value zero reset level even if the power is turned OFF.
As shown in the figure below, use the zero reset memory still set to "OFF" when using this function to perform
a zero reset at each measurement.
Example: When the level difference of a measurement target is measured
• When zero reset memory is set to "ON", the zero reset is written to EEPROM (non-volatile memory) in the Sensor
Controller each time that a zero reset is performed. The EEPROM can be written a maximum of 1,000,000 times.
Pay attention to the maximum number of writing allowed when this is set to "ON."
• Even if zero reset memory is disabled, the zero reset level will be held also when the setting is saved. In this
instance, the zero reset will also be continued after a restart.
Item
Setting item
Setting value
Description
Sensor settings
Zero reset memory
OFF (default
value)
The zero reset is canceled when the power is turned OFF.
ON
The zero reset level is held even if the power is turned OFF.
This step is
measured.
Sensor head
Zero reset
executed.
Direction of movement
Zero reset
executed.
Zero reset
executed.
Important
Содержание ZW-C1*T
Страница 17: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 16 1 2 Basic Operation Flow 17 ...
Страница 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Страница 116: ...Setting the System 114 ZW User s Manual MEMO ...
Страница 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Страница 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Страница 319: ...Index ZW User s Manual 317 Index 11 ...
Страница 321: ......