APPENDIX B ETHERNET COMMUNICATIONS
XR Series power supplies are available with an optional Ethernet interface. When
specified at time of order, an Ethernet interface module is installed to make a second
UART port available for communications. With two UART ports available, RS232 and
Ethernet, the one first receiving communications after power on is the port that is activated.
Once activated, the other UART port cannot be recognized unless there has been a period
of inactivity for 5 minutes. After this period, a new UART port can be recognized by
sending communications.
Ethernet communications can be made though the Remote Interface Software, a terminal
emulation programs like Microsoft Windows (tm) HyperTerminal Program, or through a
computer’s web page browser. In the latter case, software, programmed into the Ethernet
interface module, allows the power supply to be recognized and to communicate directly
with the computer without the need for additional programs stored in the computer.
An IVI-COM Driver is included with the installation CD and is available for download
from the Magna-Power Electronics’ web page. The driver allows the power supply to
communicate through many different programming languages. The driver handles the low-
level bus protocols simplifying automation development. It supports TCP/IP, GPIB, and
RS232 using standard VISA resource descriptors. For additional documentation and
details on using the IVI-COM driver, see the driver's included help file (MPEIVI.chm).
B.1
Ethernet Communications using the Remote Interface Software
Chapter 5.0 describes application of the Remote Interface Software which is shipped with
XR Series power supplies. This software provides the user with a quick method to operate
a Magna-Power Electronics’ power supply under computer control. The software can be
configured for a number of communication interfaces; Ethernet is included as one of these
interfaces.
To use the Remote Interface Software with optional Ethernet communications, the address
parameters in the Remote Interface Software and that encoded in the power supply must
match. The factory default for DHCP is 1 (on) allowing the IP address to be automatically
set by the DHCP server on the network. If DHCP is set to 0 (off), the power supply’s
address can be manually set. See Section 4.3.12 for setting address parameters. If DHCP
is off and the address is unknown, the device can be located by following one of the
procedures described in Section B.3.2.3.
B.2
Ethernet Communications using HyperTerminal
The following procedure outlines using Microsoft Windows (tm) HyperTerminal Program
to communicate using the Ethernet communications.
1.
Start the HyperTerminal application and when prompted select “TCP/IP
107
Содержание XR III series
Страница 1: ...OPERATING AND SERVICE MANUAL XR SERIES III DC POWER SUPPLIES...
Страница 2: ......
Страница 3: ...MAGNA POWER ELECTRONICS INC 39 ROYAL ROAD FLEMINGTON NJ 08822 February 20 2012...
Страница 4: ......
Страница 88: ...Figure 4 1 Status Byte Generation Figure 4 2 ESE and ESR Generation 76...
Страница 95: ...IEEE Standard CLS ESR ESE STB SRE IDN SAV RCL RST Notes 1 C command Q query 83...
Страница 97: ...Figure 5 1 Configuration setup Figure 5 2 GPIB communications setup 85...
Страница 99: ...Figure 5 4 Virtual Control Panel Figure 5 5 Command Panel 87...
Страница 102: ...Figure 5 7 Calibration Panel Figure 5 8 Firmware Panel 90...
Страница 103: ...Figure 5 9 Modulation Panel 91...
Страница 123: ...Figure B 1 Information Panel Figure B 2 Configure Panel 111...
Страница 124: ...Figure B 3 Reboot in Progress Panel Figure B 4 Web Control Panel 112...