Probes and Biopsy 12-9
The above is a dimension illustration of the 6CV1s probe.
You can select the probe sheath according to the actual application situation.
12.1.5 Probes Cleaning and Disinfection
After completing each examination, clean and disinfect (or sterilize) the probes as
required. When biopsy procedures have been performed, be sure to sterilize the
needle-guided bracket. Fail to do so may result in the probe and the needle-guided
bracket to becoming sources of infection. Please follow the instructions in the manual for
cleaning.
WARNING:
Never immerse the probe connector into liquid such as water
or disinfectant. Immersion may cause electrical shock or
malfunction.
CAUTION:
1. When performing cleaning and
disinfection of the probe to prevent
infection, wear sterile gloves.
2.
After disinfection, rinse the probe thoroughly with sterile
water to remove all chemical residues. Chemical residues
on the probe may be harmful to the human body.
3
No cleaning and disinfecting may result in the probe
becoming a source of infection.
NOTE:
1. After the examination, wipe off the ultrasound gel thoroughly. Otherwise,
the ultrasound gel may solidify and degrade the image quality of the
transducer.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...