12-40 Probes and Biopsy
Hydrogen Peroxide and Peroxyacetic Acid -based sterilant:
Trade Name
Chemical Name
Procedures
Minncare
®
Cold
Sterilant
22% Hydrogen
Peroxide
4.5% Peroxyacetic
Acid
Dilute the sterilant with sterilized purified
water (1:20).
Immersed time: 11 hours.
Temperature: 20
℃
-25
℃
.
Please refer to the instructions provided by
the manufacturer of the solution for details.
Before safety and performance is affected, plastic bracket NGB-007 can be sterilized by
Minncare COLD STERILANT for at least 245 times (11 hours for one time).
High-pressure steam sterilization (only applicable for metal guided-bracket)
Autoclaving (moist heat) 121
°
C for 20 minutes.
NOTE:
1. Repeated sterilization may degrade the safety and performance of the
needle-guided bracket.
2. The high-pressure steam/ immersion sterilization do not affect the bracket
dual life, and the dual life is affected by the daily application of the bracket.
Please check the appearance of the bracket before using.
12.2.10 Storage and Transportation
1. Don’t use the carrying case for storing the needle-guided bracket. If the carrying case
is used for storage, it may become a source of infection.
2. Between examinations, keep the needle-guided bracket in a sterile environment.
3. When the needle-guided bracket is sent to your MINDRAY representative for repair,
be sure to disinfect or sterilize it and keep it in the carrying case to prevent infection.
4. Sterilize the carrying case as necessary.
5. Store or transport the needle-guided bracket under the following ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 30% to 85% (no condensation)
12.2.11 Disposal
Be sure to dispose the needle-guided bracket only after sterilizing it.
Contact your MINDRAY representative when disposing of this device.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...