System Preparation 3-3
Turning on the Power
To turn on the system:
1. Press the power button in the upper right corner on the control panel
2. The system enters the work status.
3. The indicator lights on and becomes green.
To check the system after it is turned on
Press the power button in the upper right corner on the control panel to turn on the
system.
To check the system after the system is turned on:
No.
Check Item
<1>
There shall be no unusual sounds or smells indicating possible overheating.
<2>
There shall be no persistently displayed system error message.
<3>
There shall be no evident excessive noise, discontinuous, absent or black artifacts
in the B mode image.
<4>
Check if there is abnormal heat on the surface of the probe during an ultrasound
procedure.
<5>
The control panel keys and knobs are fully functional.
<6>
The exam date and time are the same as the system date and time, and are
displayed correctly.
WARNING:
1. If you use a probe giving off excessive heat, it may burn
the patient.
2. If you find anything not functioning properly, this may
indicate that the system is defective. In this case, shut
down the system immediately and contact Mindray
Customer Service Department or sales representative.
NOTE:
When you start the system or switch between probes, you will hear clicking
sounds – this is expected behavior.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...