System Maintenance 18-5
18.1.2 Checking the Probe
Visually check to confirm that there is no crack and expansion to probe head.
Visually check to confirm that there is no deterioration or desquamation to probe
cable.
Visually check to confirm that there is no bend, destroyed or falling off pins to the
connector.
18.1.3 Checking the Power Cable and Plug
Visually check to confirm that there is no wrinkles, crack or deterioration; No crack
and expansion to the surface of adapter.
Manually check to confirm that there is no looseness or rupture. The connection of
plug is reliable.
18.1.4 Checking Appearance
Check if there is any crack in the covers:
Ultrasound system covers
Probe appearance
External appearance of ECG lead
18.1.5 Backup of the System Hard Drive
To prevent deterioration or loss of data stored in the system hard drive, create a backup
copy of the hard drive data (including patient info data, preset data and etc.) at regular
intervals.
18.2 Troubleshooting
If any persistent system malfunction is experienced, e.g. an onscreen error message,
blank imaging screen, absent menus, please refer to the following table below. If the
failure cannot be eliminated, please contact Mindray Customer Service Department or
sales representative.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...