DICOM 11-7
Name
Description
Interval Time(s) Reserved feature.
Timeout
Refers to time after which the system will stop trying to establish a
connection to the service. Value: 5
-
60s.
Allow
Multiframe
If SCP supports this function, then select it.
Max Framerate
Set the frame range of transferring cin file into DCM multi-frame file.
SR Storage
Option
To enable or disable structured reporting sending.
Cine Zoom
Mode
Select the cine zoom mode during image file storage.
Compression
Mode
Select the compression mode: uncompressed, RLE, JPEG and
JPEG2000.
Compression
Ratio
Select the compression ratio of JPEG: lossless, low, medium and high.
Color Mode
Select the color mode.
Cancel
Click to cancel the parameter setting.
Update
Select an item in the service list, change the parameters in the above
area, and click [Update] to update the item in the service list.
Delete
Click to delete the selected service in the service list.
Default
Select an item in the service list, click [Default] and you can see “Y” in
the Default column.
Verify
Click to verify if the two DICOM application entities are normally
connected.
Exit
Click to exit the screen.
Tips: RLE, JPEG and JPEG2000 are not supported by all SCPs. Please refer to the
electronic file
DICOM CONFORMANCE STATEMENT
of the SCP to check whether SCP
supports or not. Do not select these compression modes if the storage server doesn’t
support them.
11.1.4.2 Print Service Setting
1. Enter the “Print” page: “[Setup]→[Network Preset]→[DICOM Service]→Print.
Select device, enter the information. For device setting, please refer to “11.1.3
DICOM Server Setting”.
Click [Add] to add the service to the Service list.
Select an item in the service list, change the parameters in the above area, and
click [Update] to update the item in the service list.
Select an item in the service list, and click [Delete] to delete the service.
Select an item in the service list, and click [Default] to set the server to be the
default service.
2. Select an item in the service list, and click [Verify] to verify the connection.
3. Click [Exit] to confirm the setting and exit the page.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...