12-6 Probes and Biopsy
12.1.3 Procedures for Operating
This section describes general procedures for operating the probe. The proper clinical
technique to be used for operating the probe should be selected on the basis of
specialized training and clinical experience.
Procedures for operating (with biopsy function):
Inspection before examination
Connection to the ultrasonic
diagnostic system
Examinations
Biopsy procedure
Wiping off the ultrasound gel
Washing the transducer with water
Draining/drying
Immersion into disinfectant
Draining/drying
Inspection after use
Storage
Disconnection to the ultrasonic
diagnostic system
Sterilization of the needle-
guided bracket
Inspection after use
Storage
Rinsing the transducer in sterile water
Removing the transducer
from disinfectant
Содержание M5 Exp