11-8 DICOM
DICOM print setting items are described as follows:
Name
Description
Device
After you set the server (s) in DICOM Server Setting, the name(s) will
appear in the drop-down list, select the name of the print server.
Service Name
Default is xxx-Print, and it can be modified.
AE Title
Application Entity title, here, it should be consistent with that of the print
server.
Port
DICOM communication port, 104 by default. Here, the port should be
consistent with that of the print server port.
Maximum
Retries
Refers to times that the system performs retrying connection, the
adjusting range is 0-9.
Interval Time(s) Reserved feature.
Timeout
Refers to time after which the system will stop trying to establish a
connection to the service.
Value: 5
-
60s, in increments of 5s, and 15 by default.
Copies
Refer to copies of printed files. You can select among 1 through 5, or
directly enter the number.
Settings
The system supports RGB (color printing) and MONOCHROME2 (black
and white printing). Please select the type the printer supports.
Film Orientation Select between LANDSCAPE and PORTRAIT.
Priority
Specify printing task priority among HIGH, MED and LOW.
Film Size
Select film size among the selections listed in the drop-down list.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...