A-2 Wireless LAN
c) Local TCP/ IP preset items are described as follows:
Name
Description
Station Name
Name of the ultrasound system.
Station Location
Location or department where the system is placed.
Current Net Adapter To select net adapter type.
DHCP/ Static
If “DHCP” is selected, IP address will be automatically obtained
from DNS server; if “Static” is selected (using static IP address),
you need to enter the IP address.
IP Address
IP address of the system.
Subnet Mask
Used to set different network segment.
Gateway
Used to set the gateway IP.
OK
Click to confirm parameter setting.
Cancel
Click to cancel parameter preset.
The IP address of the system should not be occupied by other devices in the LAN, or
failure may result.
NOTE
DO Not edit IP address in Network Preset page as the network task is executed,
otherwise it fails. View if there is any undergoing task in task manager of the
system.
3. Click the wireless LAN manager icon
in the lower right corner of the toolbar
region, and the manager dialogue box pops up:
[Connect]: select the target network and click to connect, and the manager icon
turns
after connected.
[Disconnect]: select the target network and click to disconnect.
[Refresh]: click to update the wireless network list.
[Configure]: click to enter the “Local TCP/IP Setting” screen.
[Show Detail]: click to view the connection information of the wireless network,
and select [Export] in the prompted dialogue box to save the information.
[Set Preferred Network]: click to set the selected network to be a default
connecting network, and the system will automatically connect to this network
after detecting it.
[Delete Profile]: click to delete the preferred network setting and password that
has been entered.
NOTE:
1. When connecting an encrypted network, entered password will be
remembered.
2. When there is more than one wireless router in the LAN, DO NOT disable
the DHCP function of the current router.
3. If error password has been entered when connecting an encrypted network,
click [Delete Profile] and reconnect the network.
Tips: if abnormal network performance is detected, e.g. wireless network list is not
updated in time; you may take the following measures:
1. Connect/disconnect the network adapter again.
2. Restart the router and ultrasound system if the problem is not fixed.
3. Restart the router again and wait for 5-10 min and reconnect the network again if the
problem is not fixed.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...