Probes and Biopsy 12-31
(3) Screw the pinch nut of the needle-guided bracket to ensure that the
needle-guided bracket is properly installed on the transducer.
(4) Select a suitable guiding block and push it into the groove above the angle block.
(5) Screw the block's nut to secure the block.
(6) Insert a biopsy needle with the same specification as that of the guiding block into
the guiding block hole.
NGB-024
1. Put on the sterile transducer sheath.
2. Select a proper needle-guided bracket, and match the groove with the tab of the
transducer. Mount the bracket onto the transducer.
3. Rotate the clamping nuts of the guided bracket on right and left side to fix the bracket
and the transducer. Rotate the needle-type adjusting nut to the ultimate position as
shown in the figure.
4. Hold the transducer. Press the biopsy needle to separate needle guided V-shaped
block from pressure position of the needle.
5. Put the needle into the needle guided-bracket, and the needle leans to V-shaped
block.
6. Hold the transducer, and release the pressure position of the needle. Adjust the
needle-type adjusting nut manually (following the direction of the arrow). The needle
moves smoothly at the vertical direction due to its gravity.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...