12-8 Probes and Biopsy
12.1.4 Wearing the Probe Sheath
A legally marketed probe sheath must be installed over the probe before performing
intra-cavitary and intra-operative examination. Protective barriers may be required to
minimize disease transmission. Probe sheaths are available for use with all clinical
situations where infection is a concern.
To order probe sheath, contact:
CIVCO Medical Instruments Co.
102 First Street South, Kalona, IA 52247-9589 USA Tel: 1-319-656-4447
E-mail: [email protected]
http://www.civco.com
CAUTION:
1. Be sure to cover the probe with a new (unused) probe
sheath to prevent infection during examination. If the
package of a probe sheath is open or broken, the
sterilization of the probe sheath may not be sufficient. DO
NOT use such a probe sheath.
2. The cover contains natural rubber latex and talc that can
cause allergic reactions in some individuals.
3. DO NOT use an expired probe sheath. Before using probe
sheaths, verify whether the term of validity has expired.
Method (for reference only):
1. Place an appropriate amount of gel
inside the sheath or on probe face. Poor
imaging may result if no gel is used.
2. Insert the probe into the sheath; make
sure to use proper sterile technique.
Pull cover tightly over probe face to
remove wrinkles and air bubbles,
taking care to avoid puncturing cover.
3. Secure the sheath with enclosed elastic
bands.
4. Inspect the sheath to ensure there are
no holes or tears.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...