ECG 7-1
7
ECG
The system can be configured with the optional ECG module. In that case, the ECG signal
is displayed on the image, and can be reviewed simultaneously with the image.
When the system is connected with the ECG module:
The Physio menu is available.
The ECG icon including heart-shaped symbol and heart rate symbol are
displayed on the screen.
The system supports ECG signal from ECG lead.
WARNING:
1. DO NOT use the physiological traces for diagnosis and
monitoring.
2. To avoid electric shock, the following checks shall be
performed prior to an operation:
The ECG electrode cable shall not be cracked, frayed or
show any signs of damage or strain.
The ECG electrode cable shall be correctly connected.
You should use the ECG leads provided with the ECG
module. Otherwise it may result in electric shock.
3. The ECG electrode cable must be connected to the
system first. Only after the cable is connected to the
system, can the patient be connected to the ECG
electrodes. Failure to follow these instructions may
subject the patient to electric shock.
4. DO NOT place the ECG electrodes directly in contact the
patient’s heart; otherwise it may lead to stop of the
patient’s heartbeat.
5. DO NOT apply the ECG electrodes if the voltage exceeds
15 volts. This could produce an electric shock.
6. Before using high frequency electric surgical unit, high
frequency therapeutic equipment or defibrillator, be sure
to remove the ECG electrode from the patient, in order to
prevent electric shock.
7. Conductive parts of electrodes and associated
connectors for ECG should not contact other conductive
parts including earth/grounding.
8. Frequent trampling or squeezing on the cables may
result in cable break-down or fracture.
NOTE:
When abnormality is detected in physio trace, please check if ECG leads are
properly connected with the system.
Содержание M5 Exp
Страница 1: ...M6 M6T M6 Exp M6s M6 Pro M55 M58 M5 Exp Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 12: ......
Страница 41: ...System Overview 2 11 UMT 200 UMT 300...
Страница 246: ...12 2 Probes and Biopsy V10 4B s CW5s 4CD4s P12 4s 7L4s L12 4s P7 3s L14 6Ns P4 2s CW2s...
Страница 286: ......
Страница 288: ......
Страница 336: ......
Страница 338: ......
Страница 357: ...P N 046 008768 00 V1 0...