MEDRAD
®
Mark 7 Arterion Injection System
13 - 76
Rotate head up and purge
Use the handle and the back of the Injector
Head (but not the Manual Knob or tubing) to
rotate the Head in the Purge position (upward),
and purge the system of all air. A minimum
purge of 3.5 mL is required for the system to
recognize it as a purge.
Rotate head down to arm
Use the handle and the back of the Injector
Head (but not the Manual Knob or tubing) to
rotate the Injector Head in Inject position
(downward).
Flow Rate Reduced
The system reduced the flow rate during the
injection.
Injection Complete
The injection has completed.
Disconnect Patient
An operator has selected
End Case
and
Yes
on
the confirmation popup to end the case.
Disconnect the patient.
Case Ended
Displays after an operator has selected
End
Case
and
Yes
on the confirmation popup.
Rotate Syringe and Remove
Displays after an operator has selected
End
Case
and
Yes
on the confirmation popup and a
Syringe is present.
Procedure Halt - Display touch
The injection stopped, an operator touched the
display touch screen.
Procedure Halt - Head Touch
The injection stopped, an operator pressed a
button on the Injector Head.
Procedure Halt - ISI
The injection stopped, the imaging system
interface terminated the injection.
Procedure Halt - Low Volume
The injection stopped, an insufficient volume
condition occurs after an injection during a Multi
Arm injection.
Procedure Halt - Start Switch
The injection stopped, an operator released the
start switch during a mL/s injections or pressed
the start switch a second time during a mL/m
injection.
Calibration Needed
Have the system calibrated by Bayer HealthCare
Services or qualified Services personnel.
Table 13 - 1: Sentinel Messages
Sentinel Message
Description/Resolution
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......