Arming and Injecting
11 - 71
8.
Visually confirm that all air bubbles have been removed from the syringe. Tap the pressure
jacket after filling to facilitate air removal.
9.
Secure the distal connector of the high pressure connector tubing.
10.
Turn the Manual Knob clockwise to push contrast out until all of the air bubbles have been
removed from the high pressure connector tubing. Gentle tapping at any connection point
may be needed to facilitate air removal.
11.
Use the handle and the back of the Injector Head (but not the Manual Knob or tubing) to
rotate the Injector Head to the Inject (downward) position.
12.
Secure the catheter hub in one hand and the distal rotating luer of the high pressure
connector tubing in the other hand.
13.
Advance the piston using the Manual Knob prior to connecting the high pressure connector
tubing to a catheter or other non-Bayer disposables including administration sets and
additions to administration sets, such as but not limited to bleed back control devices and
pressure transducers.
14.
Securely tighten the high pressure connector tubing connection to the catheter or other non-
Bayer disposables including administration sets and additions to administration sets, such as
but not limited to bleed back control devices and pressure transducers.
15.
Aspirate via the manual knob when connecting to the high pressure connector tubing and a
catheter or other non-Bayer disposables including administration sets and additions to
administration sets, such as but not limited to bleed back control devices and pressure
transducers to ensure all air has been removed from the fluid path.
16.
Stop aspirating with manual knob once blood is visualized in high pressure connector tubing.
17.
Verify that there is no air in the high pressure connector tubing.
a.
If there is air in the tubing, disconnect from patient, remove air, and try fluid to fluid
connection again.
18.
Once it has been established that all air has been removed, advance the Manual Knob
forward to clear blood from high pressure connector tubing.
11.5.1 Refilling Syringe with 15 mL Purge Feature Enabled
For a full explanation of the system’s purge options and configurations, refer to
Purge Feature and Choosing Configuration Options"
.
NOTE:
With 15 mL Purge turned
ON
or
OFF
, the system will require a purge in the upright
position during system power-up, syringe installation, and new case.
15 mL Purge ON
15 mL Purge OFF
≥
15 mL Plunger
Retraction
<
15 mL Plunger
Retraction
≥
15 mL Plunger
Retraction
<
15 mL Plunger
Retraction
Arming
Requirements
MUST
purge 3.5
mL or more with
Injector in upright
position
NOT REQUIRED
to
purge
MUST
purge 3.5
mL or more
ONLY
when retraction
occurs in Injector’s
upright position
NOT REQUIRED
to
purge
Auto-Fill
remains available for use
Auto-Fill
only available in Injector’s
upright position
NOTE:
Operator can still choose to
purge prior to arming, if desired, with
plunger retraction of
<
15 mL but
system will not require it
NOTE:
Operator can still choose to purge,
if desired, with plunger retraction of
<
15
mL and/or upon plunger retraction in
Injector’s downward position
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......