MEDRAD
®
Mark 7 Arterion Injection System
16 - 116
Figure 16 - 50: Floor Mount Mounted to Floor
NOTE:
The minimum bend radius for the Injector Head Extension Cable is seven inches
and five inches for the Display Control Unit Cable.
16.7 Stand Mounting Kit Installation
The Mark 7 Arterion Pedestal Mounting Kit is for use with the KMA 320 RT and the KMA 330.
NOTE:
Ensure the KMA 320 RT and KMA 330 have five (5) locking casters. If the stand
does not have five locking casters, contact service for part number 699-4645-100.
16.8 Power Unit Bracket Installation
1.
Rotate the Pedestal Arm counterclockwise until you hit the stop.
2.
Attach Power Unit Mounting Bracket Plate (A) to the Power Unit Tube Clamp (B) using two 10-
32 x 3/8 pan head Phillips screws.(C)
3.
Repeat step
4.
Attach the assembled Power Unit Brackets to the top and bottom of the Power Unit.
NOTE:
The screw head aligns with the hole in the bracket plate.
D
E
A
B
C
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......