MEDRAD
®
Mark 7 Arterion Injection System
15 - 84
3.
Clean the Syringe Heat Maintainer with a dampened cloth using soap and water.
4.
Remove the Drop Front Cover.
5.
Clean the Drop Front Cover with a soft cloth or a paper towel dampened with a cleaning
solution to remove contrast media and other contamination.
6.
Remove the Pressure Jacket.
7.
Clean the Pressure Jacket with a soft cloth or a paper towel dampened with a cleaning
solution to remove contrast media and other contamination.
Some cleaning agents react with the plastic material and may cause structural degradation.
Bayer recommends that the Pressure Jacket be washed in a solution of warm tap water (35°
– 45° C) and mild non-abrasive detergent (neutral grade low pH, enzymatic cleaner), and
then rinsed thoroughly and dried with a soft towel.
A solution of dish washing detergent and water is compatible with the Pressure Jacket. If a
germicidal cleaning agent is desired, contact the germicide manufacturer to check the
recommended dilution and compatibility with polycarbonates. If the solution is acceptable,
follow the manufacturer’s directions exactly. Do not clean the Pressure Jacket with an
automatic dishwasher. The Pressure Jacket is not dishwasher safe. Do not leave the
Pressure Jacket in germicide for extended periods of time. Do not expose the Pressure
Jacket to fluorocarbons (such as Freon), or other solvents (acetone, benzol, carbon
tetrachloride, MEK, MIBK, toluol, trichlor, and triclene). Gasses used to pressurize aerosol
cans can be damaging to the Pressure Jacket. Therefore, do not use aerosols in or around
the Pressure Jacket.
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......