Installation - System and Accessory
16 - 105
Figure 16 - 27: Wall Mount Friction Plate - Close-up
2.
Remove the wing nut (B) from the top of the pivot bolt (A).
3.
Remove the washer and spring (C).
4.
Lift the mounting plate (D) off of the mounting pins and flip over.
5.
Install the spring and washer (C) onto the mounting pin.
6.
Install wing nut (B) on mounting bolt (A) and tighten down until it stops.
7.
Install the Display Control Unit onto the wall mount bracket.
16.6 Accessory Installation
•
"Syringe Heat Maintainer Installation"
•
"Syringe Pressure Jacket Installation"
•
"Hand Switch and Foot Switch Installation"
•
•
"Display Control Unit Sterile Sheath Installation"
•
16.6.1 Syringe Heat Maintainer Installation
Figure 16 - 28: Syringe Heat Maintainer Port (A)
NOTE:
Remove and discard cap covering the port (A).
A
B
C
D
A
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......