MEDRAD
®
Mark 7 Arterion Injection System
16 - 110
cient length of the Display Control Unit Cable extending beyond the opening of the bracket to
connect to the Display Control Unit in the desired location.
NOTE:
When installing only one cable, insert the appropriate sized plug (included) into the
empty hole.
8.
Re-assemble the insert (C) using the socket head screws (D) removed in step
.
9.
Mount the bracket to the table using existing hardware if already available on table. Use the
included #8 sheet metal screws if no existing hardware is available.
Figure 16 - 38: Table Mount Mounted to Table
NOTE:
The minimum bend radius for the Injector Head Extension Cable is seven inches
and five inches for the Display Control Unit Cable.
16.6.6.1.2 Installing the Insert Only
NOTE:
Before drilling, check clearance behind the desired mounting location inside the
table. Ensure that no cables or other hardware will interfere with the installation
and function of the Table Cable Insert.
1.
Pre-drill mounting holes using the Table Mount Cable Maintenance Insert Hole Template
provided in Appendix A.
2.
Route Injector Head Extension Cable and Display Control Unit Cable through the opening on
the table.
3.
Locate the Table Cable Bracket and disassemble. Remove the four screws attaching the
insert to the bracket.
4.
Attach the top of the insert to the inside of the opening.
5.
Align the cables in the corresponding sized slots in the insert. Leave 4 to 6 inches of the
Injector Head Extension Cable extending beyond the opening of the bracket. Leave a
sufficient length of the Display Control Unit Cable extending beyond the opening of the
bracket to connect to the Display Control Unit in the desired location.
NOTE:
When installing only one cable, insert the appropriate sized plug (included) into the
empty hole.
6.
Attach the bottom of the insert to the top of the insert.
NOTE:
The minimum bend radius for the Injector Head Extension Cable is seven inches
and five inches for the Display Control Unit Cable.
16.6.6.1.3 Installing the Bracket and Panel Mount Plate
NOTE:
Before drilling, check clearance behind the desired mounting location inside the
table. Ensure that no cables or other hardware will interfere with the installation
and function of the Table Cable Bracket.
1.
Pre-drill mounting holes (if not already provided on the table) using the Table Mount Cable
Bracket Hole Template provided in Appendix A as a guide.
Mechanical Hazard - Minor or moderate patient and/ or worker injury may result.
•
Do not create a trip hazard when installing cable bracket.
CAUTION
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......