MEDRAD
®
Mark 7 Arterion Injection System
6 - 30
6.1.2 Actuals Window
The
Actuals
window displays Peak (maximum Flow Rate achieved), Delivered (actual total volume
delivered), Total Contrast (total volume delivered for the current case), and the
End Case
button. An
operator presses the
End Case
button after completing a patient procedure and before removing the
disposables to retract the syringe plunger. This button also zeroes the Total Contrast and creates a new
case entry on the
History
tab. For more information on cases, see
.
6.1.3 Sentinel Window
The Sentinel window displays system messages, such as “Rotate head down to arm”. The Sentinel
window also displays the status of the last injection.
The Sentinel window message flashes momentarily when the system cannot arm.
A list of messages is available in Chapter 13 “System Messages.”
6.2 Protocols Tab
Operators can store new protocols, recall stored protocols, and edit existing protocols on the
Protocols
tab. For information on how to manage protocols from the
Protocols
NOTE:
To set protocol parameters from the
Home
tab, see
"9.1 - Set Injection Parameters
6.3 History Tab
The
History
tab shows a list of recent injections with date and time of the injection and the
programmed parameters and actuals. Injections are grouped by case. Cases are reset when an
operator selects the
End Case
button from the
Home
tab.
For example, an operator may want to group all of the injections for one patient together. The operator
performs all of the injections for a patient. The operator then selects the
End Case
button after the last
injection has completed. The system groups all of the injections for that patient together. This group of
injections can be retrieved from the
History
tab.
6.4 Options Tab
Operators can modify system settings from the
Options
tab. A green LED displays beside the current
setting for each option. When an operator changes a setting, the system adds an asterisk next to the
option name to identify unconfirmed setting changes. To enable the new settings, select one of the
other tabs, such as
Home
. A popup displays requesting confirmation for the new value. Select
Yes
to
confirm the setting. Select
No
to go to the selected tab without saving the changes; the system reverts
to the previous settings.
Table 6 - 1: Options
Option
Description
Language
Sets the Display Control Unit display language.
Flow Rate
Select mL/m or mL/s.
Fill Volume
Determines the volume of contrast drawn into the syringe when an
operator presses the
Auto-Fill
button on the Injector Head.
Fill Speed
Determines the speed at which the injector draws contrast into the
syringe when an operator presses the
Auto-Fill
button on the Injector
Head.
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......