MEDRAD
®
Mark 7 Arterion Injection System
16 - 112
Figure 16 - 41: Mount Plate to Bracket
8.
Mount the bracket to the table using existing hardware if already available on the table. Use
the included #8 sheet metal screws if no existing hardware is available.
Figure 16 - 42: Table Mount Mounted to Table
NOTE:
The minimum bend radius for the Injector Head Extension Cable is seven inches
and five inches for the Display Control Unit CAble.
16.6.6.2 Floor Cable Bracket Installation
NOTE:
Before drilling, check clearance behind the desired mounting location on the floor.
Ensure that no cables or other hardware will interfere with the installation and
function of the Floor Cable Bracket.
1.
Pre-drill mounting holes using the template provided in Floor Mount Cable Bracket Hold
Template in Appendix A as a guide.
NOTE:
DO NOT photocopy the template as photocopying may distort the image. ALWAYS
MEASURE BEFORE DRILLING.
2.
Thread
MEDRAD Mark 7 Arterion Injection System
Injector Head Extension Cable and Display
Control Unit Cable through the floor opening.
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......