Using and Understanding the Display Control Unit Screen
6 - 31
6.4.1 Modify Options
To modify an option:
1.
Select the
Options
tab. A list of options displays. For an explanation of the options, see
2.
Select an option. A center panel displays with setting choices or a numeric keypad based on
the option.
3.
Make the appropriate changes.
4.
Select another tab to save the changes. A popup displays requiring confirmation to save the
changes or revert back to previous settings. Select
Yes
to save changes. Select
No
to revert
back to previous settings.
6.5 Help Tab
The
Help
tab provides a list of help topics.
To view a help topic
1.
Select the
Help
tab.
2.
Select the topic name.
ISI
Enables or disables Imaging System Interface (Single mL/s protocols
only).
Auto Retract
Enables or disables the piston auto retract feature. See
for more information.
Phased
Enables or disables Phased protocols.
Date/Time
Sets the system date and time and the calibration date.
Pressure Units
Sets the pressure measurement to PSI or kPa. PSI is set by default.
Head Audio Volume
Sets the audio volume level for the Injector Head.
DCU Audio Volume
Sets the audio volume level for the Display Control Unit.
Upgrade
Used by Bayer HealthCare Services or Bayer trained personnel to
activate features.
MEDRAD
®
VFlow
Enables Variable Flow Rate injections when this feature is activated
from Upgrade. Contact Bayer to enable this feature.
Audio Feedback
When VFlow
is enabled, the system provides a sound when the Hand
Controller plunger is depressed.
15 mL Purge
With this feature enabled, the system provides a choice between two
configurations (ON or OFF) of the 15 mL Purge Option. Contact Bayer
to enable this feature.
Table 6 - 1: Options
Option
Description
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......