Installation - System and Accessory
16 - 111
NOTE:
DO NOT photocopy the template as photocopying may distort the image. ALWAYS
MEASURE BEFORE DRILLING.
2.
Route Mark 7 Arterion Injection System Injector Head Extension Cable and Display Control
Unit Cable through the table opening.
3.
Locate the Table Cable Bracket (A). Remove the four screws (B) attaching the insert to the
bracket. Control Unit Cable through the table opening.
Figure 16 - 39: Table Cable Bracket Assembly
4.
Discard the screws and the insert.
5.
Align the cable connectors to the corresponding sized slots in the mount plate.
Figure 16 - 40: Attach Cables to Panel Mount Plate
NOTE:
If attaching the DCU cable, remove the cover plate (C) from the plate.
6.
Insert the cables connectors in the corresponding sized holes in the plate.
7.
Attach the Panel Mount Plate to the bracket using the four 4-40 flat head, thread cutting
screws.
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......