System Messages
13 - 77
13.3 Popup Messages
Popups display on the touch screen and require the operator to make a selection on the screen to close
the Popup.
Table 13 - 2: Popup Messages
Popup Message
Description/Resolution
WARNING - Do Not Inject Air. Have you expelled
all air from syringe/disposable?
The Check for Air popup displays when initially
arming the injector.
This message displays again after any powered
reverse piston motion or the drop front
disengages.
Confirm the disposable system is free of air, and
select
Yes
.
Ensure the patient is disconnected. Are you sure
you want to end the case?
Displays after an operator selects End Case.
Select
Yes
to acknowledge that you want to end
the case, and that the patient has been
disconnected from the system. Select
No
to
resume the case.
Purge Terminated. Check fluid path for
occlusion.
The high pressure connector tubing or catheter
may be kinked limiting the flow of contrast.
Insufficient volume. Continue with remaining
volume? (
Single Protocol
)
The programmed volume exceed the remaining
volume in the syringe.
Select
Yes
to allow the system to adjust the
protocol to use the volume remaining in the
syringe.
Select
No
to abort the arming process. Refill the
syringe with the appropriate volume and rearm
the system.
Lock released - User inactivity timeout.
The display control unit is no longer locked out.
The timeout period has elapsed.
Insufficient volume for phased protocol. Revise
protocol.
The programmed volume exceed the remaining
volume in the syringe.
Fill the syringe with the sufficient volume or
revise the protocol. Rearm the system.
Unplug or replace Syringe Heat Maintainer.
The syringe heat maintainer is non-functional.
Select
OK
to proceed with arming process.
Injector will keep syringe heat maintainer
disabled until it can determine that the fault is
corrected or the heat maintainer is removed.
System Disarmed - Fluid flow stopped. Check
for occlusion, reduce rate or increase pressure
limit.
The system was not able to achieve the
programmed flow rate.
The high pressure connector tubing or catheter
may be kinked limiting the flow of contrast.
The programmed pressure limit is insufficient.
Verify that the pressure limit matches the
pressure rating of the disposables being used.
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......