5 - 20
5.2 Pressure Limiting
The purpose of the programmed pressure limit is to protect the patient, the catheter, and any
disposable device attached to the injector.
As a general rule, set pressure limit no higher than the max pressure rating of the weakest component
in the fluid path (tubing, stopcocks, connectors, catheters, administration sets, etc.).
Max pressure rating examples for an example scenario:
•
Tubing - 1200 psi
•
Stopcock - 1050 psi
•
Catheter - 1200 psi
In this case, set the pressure limit no higher than 1050 psi because anything higher could potentially
cause the component to fail.
Typical factors to consider and how they affect Pressure:
Consider the above factors when setting pressure limit to achieve the desired injection flow rate.
Proper pressure limit setting optimizes the angiographic images.
The injector applies the minimum pressure needed to achieve the programmed flow rate. If the
pressure from the injector exceeds the programmed pressure limit, the system cannot achieve the flow
rate and a Sentinel message displays.
Pressure information can be found in the
History
tab.
5.3 System Technical Specifications
5.3.1 Input Power Requirements
100-240 VAC
50/60 Hz
1000 VA
Factors
Effects on Pressure
Decrease
Pressure
Increase
Pressure
Fluid Viscosity
Low
High
Catheter and
Tubing Length
Short
Long
Catheter ID
Large
Small
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......