Handbook OPTIFLEX 1300 C
79 OPTIFLEX
1300
C
6.
Service and maintenance
6.1
General notes
Maintenance is normally not required in standard applications. However, heavy deposits on the probe may
cause measurement deviations or faulty operation.
Warranty statement
Servicing by the customer is limited by warranty to
•
removal and replacement of the housing containing the potted signal converter. Can be detached from
the flange system under process conditions. Refer to section 6.2.
Other repairs must be done by KROHNE-authorized service staff.
6.2
Signal converter replacement
Procedure
Do not loosen flange bolts!
Disconnect the device from the power supply!
1. Disconnect all wires (refer also to Section 2.1.2: Wiring procedure).
2. Unscrew M10 socket head set screw on the housing support pillar with a 5 mm hexagon key.
3. Remove the housing by sliding it up and off the adaptor collar pin.
4. Remount the signal converter using the same procedure.
Equipment in hazardous areas
Before replacing the signal converter in hazardous locations, make absolutely sure that there is no risk of
explosion (gas-free certificate). For further information refer to the OPTIFLEX 1300 C KEMA 04ATEX 1219
X certificate and technical data in the manual.
Содержание optiplex 1300 c
Страница 1: ......
Страница 49: ...Handbook OPTIFLEX 1300 C 49 OPTIFLEX 1300 C 1 Volume measurement of liquids...
Страница 50: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 50 2 Volume measurement of solids...
Страница 51: ...Handbook OPTIFLEX 1300 C 51 OPTIFLEX 1300 C 3 Mass measurement of liquids or solids...
Страница 62: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 62 Finding the advanced setup menu...
Страница 95: ...Handbook OPTIFLEX 1300 C 95 OPTIFLEX 1300 C Note...