Handbook OPTIFLEX 1300 C
OPTIFLEX
1300
C
20
Tank roof deformation
Ensure that the tank roof supports loads up to 3.5 metric tones or 7700 lbs.
1.3
Typical installations
1.3.1
General notes
Electromagnetic compatibility
The OPTIFLEX 1300 C conforms to European EMC standard EN 61326-1 A1+A2 and fulfills Immunity and
Emissions requirements for industrial environments. It can be installed on open-air tanks and non-metallic
(such as plastic) vessels.
Note:
When the probe is enclosed in a metal vessel or when the device is fitted with a coaxial probe, it conforms
to residential class immunity and emissions requirements.
All installations should respect the instructions given in section 1.2: Location.
Installations with difficult measurement conditions
Agitators can cause foam, turbulent product surfaces and probe entanglement.
Mounting a device in a bypass chamber or stilling well or a device with a coaxial probe are ideal solutions
for these conditions.
For further information, refer to section 1.3.3: Typical installations, stilling wells and section 1.3.4: Typical
installations, bypass chambers.
Ambient temperature
The ambient temperature limits of the device are given below.
Содержание optiplex 1300 c
Страница 1: ......
Страница 49: ...Handbook OPTIFLEX 1300 C 49 OPTIFLEX 1300 C 1 Volume measurement of liquids...
Страница 50: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 50 2 Volume measurement of solids...
Страница 51: ...Handbook OPTIFLEX 1300 C 51 OPTIFLEX 1300 C 3 Mass measurement of liquids or solids...
Страница 62: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 62 Finding the advanced setup menu...
Страница 95: ...Handbook OPTIFLEX 1300 C 95 OPTIFLEX 1300 C Note...