Handbook OPTIFLEX 1300 C
OPTIFLEX
1300
C
16
Installing two devices on the tank
Single or double probes:
Install at least 1 meter or 3.3 feet from each other to avoid interference from neighboring electromagnetic
fields.
Coaxial probes
No minimum distancing restrictions.
Two devices fitted close together: single cable probes (left) and coaxial probes (right)
1.2.2
Liquid applications
General notes
Follow these instructions and those given in section 1.2.1: location, all applications.
Turbulent products
If liquid products are agitated or turbulent, probe bending can occur.
Avoid probe bending!
Risk of measurement error!
Always install the meter as far away as possible from the source of turbulence.
Anchor the probe if necessary to ensure it stays straight and vertical.
See the following table.
Содержание optiplex 1300 c
Страница 1: ......
Страница 49: ...Handbook OPTIFLEX 1300 C 49 OPTIFLEX 1300 C 1 Volume measurement of liquids...
Страница 50: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 50 2 Volume measurement of solids...
Страница 51: ...Handbook OPTIFLEX 1300 C 51 OPTIFLEX 1300 C 3 Mass measurement of liquids or solids...
Страница 62: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 62 Finding the advanced setup menu...
Страница 95: ...Handbook OPTIFLEX 1300 C 95 OPTIFLEX 1300 C Note...