PLZ-4WH
43
2
Basi
c Funct
ions
Soft Start
You can use the soft start function to raise the input current gradually. The soft start function is
valid in constant current mode (CC mode). Soft start is useful when:
• The load is turned on at the same time that voltage is applied to the load input terminals.
• When the load is on and voltage is applied to it after a period of no input (0 V).
You can reduce the amount of output voltage distortion that occurs when the DUT starts by
setting an appropriate soft start time.
To set the soft start time (1 ms, 2 ms, 5 ms, 10 ms, 20 ms, 50 ms, 100 ms, or 200 ms), from
the menu, select “1. Setup” > “1. Function” > “Soft Start.”
Time
0
Time
0
When the soft start time is too fast
When the soft start time is appropriate
Rise time of the DUT
Rise time of the PLZ-4WH
Input current
Input current
See
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...