124
PLZ-4WH
Ending Calibration
11
Press or to set the cursor position of DAC REF, and then turn the
rotary knob so that the input voltage is within ±0.05 % of the value
corresponding to 100 % of the full scale value (see the table of CV mode
settings).
12
Press ENTER.
The load turns off automatically.
The calibration of the low range voltage values is complete.
Calibrating the high range
Calibrate the High range in the same way that you calibrated the Low range. In step 4 , select
“5. CV 650 V.” Set the voltage of the power supply to 660 V or more.
The calibration of CV mode is complete after you calibrate the High range
.
Ending Calibration
1
Press PREV (SHIFT+ ).
The screen that was displayed before the calibration screen appears.
2
Press MENU (SHIFT+SET/VSET).
The screen that was displayed before the menu appears.
When you press ENTER or NEXT (SHIFT+ ) in the gain calibration screen, the calibration
data is written to the internal memory.
If you only want to check the calibration data, be sure to press PREV (SHIFT+ ) or MENU
(SHIFT+SET/VSET) to exit the gain calibration screen.
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...