102
PLZ-4WH
Monitor Signal Output
Trigger signal output
The trigger signal can be used as a synchronization signal for monitoring the waveform of the
switching operation on an oscilloscope. It can also be used to synchronize with an external
device during sequence execution.
The trigger signal is output from the TRIG OUT connector (BNC connector) on the PLZ-4WH
front panel.
The trigger signal output voltage is approximately 4.5 V. The pulse width is 2 µs or more. The
output impedance is approximately 1 k
Ω
. The TRIG OUT connector is connected to the
chassis electric potential. It is isolated from A COM.
The trigger signal output is generated
under the following conditions.
• During switching operation.
• When a step in which trigger output
has been specified is executed
during sequence operation.
Current monitor output
The signal is output from the I MON OUT connector on the PLZ-4WH front panel and across
pins 2 and 3 (pin 8 is common) of the EXT CONT connector.
■
I MON OUT connector (BNC) on the PLZ-4WH front panel
The common is connected to the chassis electric
potential. It is isolated from A COM. For current ranges
H and L, 1 V corresponds to the full scale current; for
current range M, 0.1 V corresponds to the full scale
current. The maximum output current is 5 mA.
■
Across pins 2 and 3 of the EXT CONT connector
The common is connected to A COM. For current ranges H and L, 10 V corresponds to the
full scale current; for current range M, 1 V corresponds to the full scale current. The output
impedance is 1 k
Ω
.
1 kΩ
Approx. 2 μs
Approx. 4.5 V
0 V
TRIG OUT
Front panel
CAUTION
To avoid damaging the PLZ-4WH, observe the following precautions. Pin 3 of the EXT
CONT connector is connected the negative load input terminal. Make sure that the wire of
pin 3 does not touch any of the other pins.
Monitor output
V
H/L
V
M
0
Range
H/L M
0
Current
Full scale of the range
V
H/L
V
M
I MON OUT connector
(BNC)
10 V
1 V
Across pins 2 and 3 of the
EXT CONT connector
10 V
1 V
I MON OUT
Front panel
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...