PLZ-4WH
105
5
Ext
e
rn
al
Cont
rol
an
d Par
a
llel
Oper
at
ion
Connections and Settings during Parallel Operation Using the Same Type of Electronic Loads
Use signal wires and load wires to connect the electronic loads. The optional flat cable
(PC01-PLZ-4W or PC02-PLZ-4W) is required to connect signal wires. You can connect up to
4 slave units.
A protective socket is attached to the PARALLEL connector when the PLZ-4WH is in the
factory default state. Store this protective socket in a safe place so that it can be used during
independent operation. If it is damaged or lost, contact your Kikusui agent or distributor.
Be sure to turn off the PLZ-4WH’s power switch before you attach or remove sockets and
cables.
When you remove a socket or cable, first release the lock lever on each side of the socket or
cable, and then pull from the socket or cable.
When considering the load wires that you will connect to the PLZ-4WH, take into account the
current that you will be using, and use load wires that are as short as possible and have
sufficient thickness. We recommend that you use busbars.
Improper use of load wires may lead to operational instability. Wire the load wires so that they
are as far away from the flat cables as possible.
See
WARNING
Improper use of load wires may lead to fire. Use load wiring with a diameter that is
appropriate for the amount of current being used and with sturdy, incombustible or
flame-resistant insulation.
CAUTION
To avoid damaging the product, observe the following precautions.
• When you are performing parallel operation, be sure to use the load input terminal on the
rear panel. Do not connect other electronic loads to the load input terminals on the front
panel.
• Do not connect the PARALLEL connector’s IN and OUT terminals incorrectly.
• Be sure to specify the same number of units as are actually connected. If the wrong
number of units is specified or if a slave unit is not turned on during parallel operation, the
actual current that flows will differ from the setting and the displayed current.
No. of Slave Units Maximum Current / Maximum Power
PLZ164WH
PLZ334WH
PLZ1004WH
1
16.5 A / 330 W
33 A / 660 W
100 A / 2 000 W
2
24.75 A / 495 W
49.5 A / 990 W
150 A / 3 000 W
3
33 A / 660 W
66 A / 1 320 W
200 A / 4 000 W
4
41.25 A / 825 W
82.5 A / 1 650 W
250 A / 5 000 W
Protective socket for J1
[84-49-0071]
WARNING
To avoid electric shock, observe the following precaution. The PARALLEL connector
contains pins that have the same potential as the output terminal. When the
PARALLEL connector is not in use, be sure to attach the protective socket that comes
with the PLZ-4WH.
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...