PLZ-4WH
119
Calibration
7
M
a
in
tenance
■
Calibration numbers
The numbers that appear in the calibration screen are the calibration numbers.
Alarm
If an alarm occurs during calibration, the PLZ-4WH produces an alarm sound, and the load
turns off. After you remove the cause of the alarm and press ENTER, the alarm sound stops,
and the PLZ-4WH returns to the Calibration screen.
Start calibration again from the calibration number at which the alarm occurred.
Preparation
Before you calibrate the PLZ-4WH, leave it on for 30 minutes or more to warm it up.Warming
up reduces measurement error caused by initial drift. Keep the ambient temperature at 23 ±
5
°
C.
Equipment used
Connect the equipment as shown below. Select an appropriate shunt resistor according to
type of calibration that you are performing.
1.
CC
(
Low
)
2. CC
(
Mid
)
4. CV 65V
5. CV 650V
PREV
NEXT
3. CC
(
Hi
g
h
)
Calibration
Calibration number
Name
Required
Accuracy
Required Rating
PLZ164WH
PLZ334WH
PLZ1004WH
DC voltmeter
Within 0.02 %
Measurement voltage range:
0 V to 700 V
Shunt resistor
1
1.
In place of a shunt resistor, you can use an ammeter.
0.1 %
For 0.1 A
For 1 A
For 50 A
For 0.2 A
For 2 A
For 20 A
For 0.5 A
For 5 A
For 50 A
Regulated DC power supply
(constant-voltage power
supply)
-
Voltage: 5 V
Current: 8.25 A
Voltage: 5 V
Current: 16.5 A
Voltage: 5 V
Current: 50 A
Regulated DC power supply
(constant current power
supply)
-
Voltage: 700 V
Current: 50 mA
Regulated DC
power supply
Voltmeter
(DVM)
Voltmeter
(DVM)
PLZ-4WH
series
+S
-S
+
-
I
1
V
1
Shunt resistor or ammeter
This is unnecessary if you
are using an ammeter.
+
-
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...